Determination of Transdermal Rate of Metallic Microneedle Array through an Impedance Measurements-Based Numerical Check Screening Algorithm | |
Jingshan Mo1; Junqing Liu2; Shuang Huang1; Baoming Liang1; Xinshuo Huang1; Cheng Yang1; Meiwan Chen3; Jing Liu4; Tong Zhang5,6; Xi Xie1,4; Jun Guo2; Fanmao Liu4; Hui Jiuan Chen1 | |
2022-04-30 | |
Source Publication | Micromachines |
ISSN | 2072-666X |
Volume | 13Issue:5Pages:718 |
Abstract | Microneedle systems have been widely used in health monitoring, painless drug delivery, and medical cosmetology. Although many studies on microneedle materials, structures, and applications have been conducted, the applications of microneedles often suffered from issues of inconsistent penetration rates due to the complication of skin-microneedle interface. In this study, we demonstrated a methodology of determination of transdermal rate of metallic microneedle array through impedance measurements-based numerical check screening algorithm. Metallic sheet microneedle array sensors with different sizes were fabricated to evaluate different transdermal rates. In vitro sensing of hydrogen peroxide confirmed the effect of transdermal rate on the sensing outcomes. An FEM simulation model of a microneedle array revealed the monotonous relation between the transdermal state and test current. Accordingly, two methods were primely derived to calculate the transdermal rate from the test current. First, an exact logic method provided the number of unpenetrated tips per sheet, but it required more rigorous testing results. Second, a fuzzy logic method provided an approximate transdermal rate on adjacent areas, being more applicable and robust to errors. Real-time transdermal rate estimation may be essential for improving the performance of microneedle systems, and this study provides various fundaments toward that goal. |
Keyword | Comsol Impedance Measurement Microneedle Transdermal Rate |
DOI | 10.3390/mi13050718 |
URL | View the original |
Indexed By | SCIE |
Language | 英語English |
WOS Research Area | Chemistry ; Science & Technology - Other Topics ; Instruments & Instrumentation ; Physics |
WOS Subject | Chemistry, Analytical;nanoscience & Nanotechnology;instruments & Instrumentation;physics, Applied |
WOS ID | WOS:000803246900001 |
Publisher | MDPI,ST ALBAN-ANLAGE 66, CH-4052 BASEL, SWITZERLAND |
Scopus ID | 2-s2.0-85129945905 |
Fulltext Access | |
Citation statistics | |
Document Type | Journal article |
Collection | Institute of Chinese Medical Sciences THE STATE KEY LABORATORY OF QUALITY RESEARCH IN CHINESE MEDICINE (UNIVERSITY OF MACAU) |
Corresponding Author | Jun Guo; Fanmao Liu; Hui Jiuan Chen |
Affiliation | 1.School of Electronics and Information Technology, State Key Laboratory of Optoelectronic Materials and Technologies, Guangdong Province Key Laboratory of Display Material and Technology, Sun Yat-Sen University, Guangzhou, 510006, China 2.Department of Cardiology, The First Affiliated Hospital of Jinan University, Guangzhou, 510630, China 3.State Key Laboratory of Quality Research in Chinese Medicine, Institute of Chinese Medical Sciences, University of Macau, Macau, 999078, China 4.The First Affiliated Hospital of Sun Yat-Sen University, Guangzhou, 510080, China 5.School of Computer Science and Engineering, South China University of Technology, Guangzhou, 510006, China 6.Pazhou Lab, Guangzhou, 510335, China |
Recommended Citation GB/T 7714 | Jingshan Mo,Junqing Liu,Shuang Huang,et al. Determination of Transdermal Rate of Metallic Microneedle Array through an Impedance Measurements-Based Numerical Check Screening Algorithm[J]. Micromachines, 2022, 13(5), 718. |
APA | Jingshan Mo., Junqing Liu., Shuang Huang., Baoming Liang., Xinshuo Huang., Cheng Yang., Meiwan Chen., Jing Liu., Tong Zhang., Xi Xie., Jun Guo., Fanmao Liu., & Hui Jiuan Chen (2022). Determination of Transdermal Rate of Metallic Microneedle Array through an Impedance Measurements-Based Numerical Check Screening Algorithm. Micromachines, 13(5), 718. |
MLA | Jingshan Mo,et al."Determination of Transdermal Rate of Metallic Microneedle Array through an Impedance Measurements-Based Numerical Check Screening Algorithm".Micromachines 13.5(2022):718. |
Files in This Item: | There are no files associated with this item. |
Items in the repository are protected by copyright, with all rights reserved, unless otherwise indicated.
Edit Comment