Residential College | false |
Status | 已發表Published |
Development of a Photoelectric Adjustment System with Extended Range for Fluorescence Immunochromatographic Assay Strip Readers | |
Wu, Huihuang1,2; Gao, Yueming1,2; Yang, Jiejie1,2; Vai, Mangi2,4,5; Du, Min1,3; Pun, Siohang4 | |
2021-06-01 | |
Source Publication | IEEE Photonics Journal |
ISSN | 1943-0655 |
Volume | 13Issue:3 |
Abstract | Fluorescence immunochromatographic assay (FICA) is a quantitative detection technique widely used in clinical diagnosis, environmental monitoring, and food safety. To deal with the limited applications caused by insufficient detection range, this paper proposes a photoelectric adjustment system suitable for FICA strip readers to expand its detection range. The photoelectric adjustment system is proposed based on the relationship between the excitation light intensity and the fluorescence intensity, which provides the optimal excitation light intensity and a stable baseline amplitude for the FICA strip reader. To verify the proposed system, it was applied to the strip reader we had previously developed. The results show that the linear detection range of the FICA strip reader was extended from 1.95-256 μg/mL to 1-1024 μg/mL after using the proposed method. At the same time, the accuracy of the FICA strip reader does not deteriorate and is in good comparisons with the conventional ESEQuant Lateral Flow Reader (with R2 > 0.9987). Therefore, the proposed photoelectric adjustment method and system can improve the detection range of the strip reader or other similar devices. |
Keyword | Fluorescence Immunochromatographic Assay Photoelectric Adjustment Light Intensity Detection Range |
DOI | 10.1109/JPHOT.2021.3075900 |
URL | View the original |
Indexed By | SCIE |
Language | 英語English |
WOS Research Area | Engineering ; Optics ; Physics |
WOS Subject | Engineering, Electrical & Electronic ; Optics ; Physics, Applied |
WOS ID | WOS:000652785300001 |
Scopus ID | 2-s2.0-85105076372 |
Fulltext Access | |
Citation statistics | |
Document Type | Journal article |
Collection | DEPARTMENT OF ELECTRICAL AND COMPUTER ENGINEERING THE STATE KEY LABORATORY OF ANALOG AND MIXED-SIGNAL VLSI (UNIVERSITY OF MACAU) |
Corresponding Author | Gao, Yueming |
Affiliation | 1.College of Physics and Information Engineering, Fuzhou University, Fuzhou, 350108, China 2.Key Lab of Medical Instrumentation and Pharmaceutical Technology of Fujian Province, Fuzhou, 350108, China 3.Key Lab of Eco-Industrial Green Technology of Fujian Province, Nanping, 354300, China 4.State Key Laboratory of Analog and Mixed-Signal VLSI, University of Macau, 999078, Macao 5.Department of Electrical and Computer Engineering, Faculty of Science and Technology, University of Macau, 999078, Macao |
Recommended Citation GB/T 7714 | Wu, Huihuang,Gao, Yueming,Yang, Jiejie,et al. Development of a Photoelectric Adjustment System with Extended Range for Fluorescence Immunochromatographic Assay Strip Readers[J]. IEEE Photonics Journal, 2021, 13(3). |
APA | Wu, Huihuang., Gao, Yueming., Yang, Jiejie., Vai, Mangi., Du, Min., & Pun, Siohang (2021). Development of a Photoelectric Adjustment System with Extended Range for Fluorescence Immunochromatographic Assay Strip Readers. IEEE Photonics Journal, 13(3). |
MLA | Wu, Huihuang,et al."Development of a Photoelectric Adjustment System with Extended Range for Fluorescence Immunochromatographic Assay Strip Readers".IEEE Photonics Journal 13.3(2021). |
Files in This Item: | There are no files associated with this item. |
Items in the repository are protected by copyright, with all rights reserved, unless otherwise indicated.
Edit Comment