Abstract | BACKGROUND:Respiratory gating reduces motion blurring in cardiac SPECT. Here we aim to evaluate the performance of three respiratory gating strategies using a population of digital phantoms with known truth and clinical data. METHODS:We analytically simulated 60 projections for 10 XCAT phantoms with 99mTc-sestamibi distributions using three gating schemes: equal amplitude gating (AG), equal count gating (CG), and equal time gating (TG). Clinical list-mode data for 10 patients who underwent 99mTc-sestamibi scans were also processed using the 3 gating schemes. Reconstructed images in each gate were registered to a reference gate, averaged and reoriented to generate the polar plots. For simulations, image noise, relative difference (RD) of averaged count for each of the 17 segment, and relative defect size difference (RSD) were analyzed. For clinical data, image intensity profile and FWHM were measured across the left ventricle wall. RESULTS:For simulations, AG and CG methods showed significantly lower RD and RSD compared to TG, while noise variation was more non-uniform through different gates for AG. In the clinical study, AG and CG had smaller FWHM than TG. CONCLUSIONS:AG and CG methods show better performance for motion reduction and are recommended for clinical respiratory gating SPECT implementation. |
Keyword | Cardiac Perfusion Simulation Spect/ct Respiratory Gating |
DOI | 10.1007/s12350-018-1392-7 |
Language | 英語English |
WOS ID | WOS:000529014700039 |
Scopus ID | 2-s2.0-85051860976 |
Fulltext Access | |
Citation statistics | |
Document Type | Conference paper |
Collection | Faculty of Health Sciences |
Affiliation | 1.Biomedical Imaging Laboratory (BIG), Department of Electrical and Computer Engineering, Faculty of Science and Technology, University of Macau, Taipa, Macau SAR, China. 2.Department of Radiology, University of Massachusetts Medical School, Worcester, USA. 3.The Russell H Morgan Department of Radiology and Radiological Science, Johns Hopkins University, Baltimore, MD, USA. 4.Radiopharmaceutical Imaging and Dosimetry (RAPID), LLC, Baltimore, MD, USA. 5.Faculty of Health Sciences, University of Macau, Taipa, Macau SAR, China. |
First Author Affilication | Faculty of Science and Technology |
Recommended Citation GB/T 7714 | Zhang D,Pretorius PH,Ghaly M,et al. Evaluation of different respiratory gating schemes for cardiac SPECT[C], 2018. |
APA | Zhang D., Pretorius PH., Ghaly M., Zhang Q., King MA., & Mok GSP (2018). Evaluation of different respiratory gating schemes for cardiac SPECT. . |
Files in This Item: | There are no files associated with this item. |
Items in the repository are protected by copyright, with all rights reserved, unless otherwise indicated.
Edit Comment