UM  > Faculty of Health Sciences
Residential Collegefalse
Status已發表Published
Evaluation of different respiratory gating schemes for cardiac SPECT
Zhang D1,2; Pretorius PH2; Ghaly M3,4; Zhang Q1; King MA2; Mok GSP1,2,5
2018
Conference Name2017 IEEE Nuclear Science Symposium and Medical Imaging Conference
Conference DateOctober 21 – 28, 2017
Conference PlaceAtlanta, GA, USA
Abstract

BACKGROUND:

Respiratory gating reduces motion blurring in cardiac SPECT. Here we aim to evaluate the performance of three respiratory gating strategies using a population of digital phantoms with known truth and clinical data.

METHODS:

We analytically simulated 60 projections for 10 XCAT phantoms with 99mTc-sestamibi distributions using three gating schemes: equal amplitude gating (AG), equal count gating (CG), and equal time gating (TG). Clinical list-mode data for 10 patients who underwent 99mTc-sestamibi scans were also processed using the 3 gating schemes. Reconstructed images in each gate were registered to a reference gate, averaged and reoriented to generate the polar plots. For simulations, image noise, relative difference (RD) of averaged count for each of the 17 segment, and relative defect size difference (RSD) were analyzed. For clinical data, image intensity profile and FWHM were measured across the left ventricle wall.

RESULTS:

For simulations, AG and CG methods showed significantly lower RD and RSD compared to TG, while noise variation was more non-uniform through different gates for AG. In the clinical study, AG and CG had smaller FWHM than TG.

CONCLUSIONS:

AG and CG methods show better performance for motion reduction and are recommended for clinical respiratory gating SPECT implementation.

KeywordCardiac Perfusion Simulation Spect/ct Respiratory Gating
DOI10.1007/s12350-018-1392-7
Language英語English
WOS IDWOS:000529014700039
Scopus ID2-s2.0-85051860976
Fulltext Access
Citation statistics
Document TypeConference paper
CollectionFaculty of Health Sciences
Affiliation1.Biomedical Imaging Laboratory (BIG), Department of Electrical and Computer Engineering, Faculty of Science and Technology, University of Macau, Taipa, Macau SAR, China.
2.Department of Radiology, University of Massachusetts Medical School, Worcester, USA.
3.The Russell H Morgan Department of Radiology and Radiological Science, Johns Hopkins University, Baltimore, MD, USA.
4.Radiopharmaceutical Imaging and Dosimetry (RAPID), LLC, Baltimore, MD, USA.
5.Faculty of Health Sciences, University of Macau, Taipa, Macau SAR, China.
First Author AffilicationFaculty of Science and Technology
Recommended Citation
GB/T 7714
Zhang D,Pretorius PH,Ghaly M,et al. Evaluation of different respiratory gating schemes for cardiac SPECT[C], 2018.
APA Zhang D., Pretorius PH., Ghaly M., Zhang Q., King MA., & Mok GSP (2018). Evaluation of different respiratory gating schemes for cardiac SPECT. .
Files in This Item:
There are no files associated with this item.
Related Services
Recommend this item
Bookmark
Usage statistics
Export to Endnote
Google Scholar
Similar articles in Google Scholar
[Zhang D]'s Articles
[Pretorius PH]'s Articles
[Ghaly M]'s Articles
Baidu academic
Similar articles in Baidu academic
[Zhang D]'s Articles
[Pretorius PH]'s Articles
[Ghaly M]'s Articles
Bing Scholar
Similar articles in Bing Scholar
[Zhang D]'s Articles
[Pretorius PH]'s Articles
[Ghaly M]'s Articles
Terms of Use
No data!
Social Bookmark/Share
All comments (0)
No comment.
 

Items in the repository are protected by copyright, with all rights reserved, unless otherwise indicated.